Mechanical Engineering Faculty Research

Title

Effects of Particle's Off-Axis Position, Shape, Orientation and Entry Position on Resistance Changes of Micro Coulter Counting Devices

Document Type

Article

Publication Date

Spring 4-4-2011

Abstract

With the recent advance in micro/nano-fabrication technology, micro Coulter counters have been widely used in detecting and characterizing micro- and nanoscale objects. In this paper, the electrical resistance change during translocation of a non-conducting particle through a channel is studied numerically. The numerical results are validated by proven analytical results available in the literature. The effects of particle's off-axis position, shape and orientation, and entry position are studied for particles with a large dynamic range. From the numerical results, a new fitted correlation is proposed that can accurately predict the resistance change caused by off-axis spherical particles regardless of their size. The shape and orientation effects of the electrical resistance change are studied by changing the axis ratio of spheroid particles and their orientation angles. Results show that a particle's shape and orientation have a significant influence on the resistance change. Simulation of an entry effect indicates that a particle starts to induce a resistance change before it enters the channel and still causes a resistance change even after the particle exits the channel completely. This study will offer some guidelines in designing and implementing Coulter counting devices and experiments, and provide insights into explaining experimental results.

Publication Title

Measurement Science and Technology

Volume

22

Issue

4

First Page

045804

Last Page

045804

Share

COinS